Fast, safe, dependable and convenient steam sterilization for liquids, media, instruments, glassware, clothing and waste
- 7”, 9”, 10” and 15” diameter chambers
- Standard drying cycle and slow exhaust
- Double locking safety door
- No special plumbing or venting required
- Available in both mechanical and electronic models, some with printer
Environmental Lab Model, Thomas number 1250K35, meets all of the following EPA Guidelines:
- Temperature gauge with sensor on the exhaust
- Operational safety valve
- Depressurizes slowly to ensure media does not boil over
- Quality Control records date, contents, sterilization time and temperature for each cycle
Autoclaves 1250K02, 1250K04, 1250K38 and 1250K47 are mechanical control units with temperature range of 100-134° C, dial thermometers and analog controls; all others are electronic units and feature a 105°-137° C temperature range, microprocessor control and digital display. Either pre-set programs or user adjustable temperature, time and pressure settings are digitally displayed. Self-contained water supply controls purity of sterilizing steam. Audible and visual alarms indicate sterilization cycle interruption. A low water cut-off protects heating elements. Chamber is constructed of high grade 316L stainless steel. Autoclaves are UL listed.