The PAL-SALT allows for quick and reliable salt concentration testing through conductivity. Since the PAL-SALT is reagent and chemical free, there is no hazardous waste to dispose of, making the unit safe to use on the product line. Simply dilute 10 grams of sample with 90 grams of water, place two drops of sample onto the prism, press the “START” key and receive a reading within 3 seconds (improved efficiency is guaranteed when used together with conventional titration method). The unit features a digital display, eliminating any varied readings caused by user interpretation. Due to the IP65 rating, simply clean the prism under running water. The body is made of tough ABS resin, making the PAL-SALT durable and able to withstand impact. No calibration liquids are required; the unit is zero set with air. Clean the prism and press “START” at the beginning of each day.
- Fast reliable testing in 3 seconds
- No reagents or chemicals required
- Easy air calibration
- Compact and portable device
Measurement Range: 0.00 to 10.0%(g/100g)
Resolution: 0.01%for salt concentration of 0.00 to 2.99%; 0.1%for salt concentration of 3.0 to 10.0%
Measurement accuracy: Displayed value ±0.05%(for salt concentration of 0.00 to 0.99%)
Relative precision: ±5% (for salt concentration of1.00 to 10.0%)
Measurement Temperature: 5 to 100?